So, Beam Guide, we are currently using on the majority of our patients, and we have it on all of our systems. I’m just going to talk about the benefits for the different sites. What Beam Guide is, it’s an additional setup tool. So alongside Postural Video, your deltas, and your deformation, you also have Beam Guide. And what Beam Guide does is it gives you a real-time beam projection. I’m sure we’ve all heard of DoseRT, which gives you what you’ve treated because it collects the data after treatment using Cherenkov imaging. What Beam Guide does, it gives you a visual representation of where you’re going to treat before you deliver the radiation.

The real-time projection is a visualization of planned radiation fields. You can either have individual fields, a combination, or a VMAT arc, and you can switch them on and off as you need to. What we’ve found is, for the breast patients, we’ve tended to use all the beams together because it gives you your entry and exit for the whole of your treatment, so you can see what’s going to be delivered through all the static beams. As I say, it helps you detect and prevent errors before the radiation is delivered. Because it’s a live patient surface, the white outline is where your beam should be, and the orange is actually your patient. So what you’re really looking to do is to get that orange projection within that white outline, and then you know you’re treating in the correct area. It’s a real-time projection, and they’re available both for treatment patient setup and during treatment to monitor.

One of our radiographers coined it as having tattoos without tattoos, which I thought was a quite good explanation. So if you get your deltas all into position and you find it looks like your orange is slightly out of the white—say if it was over onto the contralateral breast—as you roll your patient, the orange moves within the white outline. It’s like when you used to roll your tattoo towards your beam edge. So that’s a good phrase that we’ve coined: having tattoos without tattoos.

One of the major benefits of this is, for anybody that’s got AlignRT already, you can get your Postural Video on, you can get your deltas on, and you can check your splash on your fields when you’re setting up your patient. But sometimes what you may not see is if your patient has dropped their chin into the field, especially when you go out of the room. What the system allows you to do, because you are actually seeing the beam and the dose that you deliver into patients, if the patient dropped their chin, even though your deltas and your Postural Video are still, then you would be able to see that you are treating the chin. Similarly, you would be able to see splash onto the contralateral breast, and you would be able to see splash into the armpit if you’re not treating that area.

When we brought the system in, one of the radiographers said to me, “What does it do?” And I said, “Well, we can remove the skin views.” And they were on board straight away because it’s quite faffy going between the Mosaiq screen, finding your screen with your skin views, then going back over to the patient and checking. Also it’s a remote image—you’re looking at something on a screen and then walking back to your patient, and then walking back to the screen and saying, “Is that correct? Do we match our skin views?” Whereas with the Beam Guide, you are actually seeing a real-time visualization of those beams on the patient. So it can speed up your workflow because you’re not having to go in between.

We have noticed that it can aid us with trickier setups. We did have a patient that was scanned with bolus, so they didn’t have a non-bolus setup surface, and they were having problems setting the patient up every day. Halfway through treatment, we brought in the Beam Guide onto that machine, and we found that we could set that patient up a lot easier using the Beam Guide, even though we didn’t have a surface without bolus on. Now, I can’t get to the bottom of why that made that setup easier, but it did make the setup easier.

Also, one of the other advantages that we did find is that an FCL came out a while ago about treating through the join of the bed. What we get our radiographers to do, especially for our pelvis patients and our thoraxes, we will ask them to take the gantry underneath the bed and center the field to make sure that at no point would any of our beams be passing through that join in the bed. What the radiographers have found is that when they’ve been setting up any lower thorax patients or pelvis patients, because you are seeing the whole VMAT arc and you can see it on the bed, you can see straight away if you would be treating through the join of the bed. So you can rectify that before you start treating the patient. But then similarly, again, you’re not having to add on that extra time of taking the gantry underneath and centering the field to check if that’s going to be an issue.

I’m just going to play a quick video of how Beam Guide works. I was going to do a video with one of our patients, but I couldn’t get permission from anybody, so I’m just going to show you this now. I’m just going to get it to the right place. There we go, I think we’re about there. Can you see that? Lovely.

As you can see on this live feed, you’ve got the white outline, and then you’ve got the projection of the beam. This is on the video tile of the AlignRT screen, and you can still flick between your three treatment cameras. So as you can see with this patient at the moment, using the postural video, the deltas, and the Beam Guide to get the patient into position, as you can see, the orange falls within the white.

Then there’s another visual representation here. You can see that the Postural Video is good, you can see the deltas are good, but the patient’s dropped their chin in. So had we not had the Beam Guide, you may not have seen that the patient’s chin was getting radiation dose.

So this was just back to the different benefits that we’ve seen for the different treatment sites. For our breasts, you can check splash on the chin, splash onto the contralateral breast or the arm, and also bolus misplacement. Because you’re able to see the beam when you put the bolus on, you can check that your whole beam is covered by the bolus.

For our DIBH workflow, you do see the beam on both the free-breathing surface and the breath-hold surface. That is worked out by the system so you can check your patient is in the right position. Although it wouldn’t be the treatment beam on the free-breathing surface, you can use all those tools to be able to check that you’re in the right position, and then when you switch over to the breath-hold surface, you then do have the beam on the breath-hold surface to check you’re in the right position.

For our pelvis and thorax, as mentioned earlier, you can visualize if you’re treating through the join of the bed, and it also helps with our supine positioning, especially on patients that would have very sloped chests where maybe the ROI can’t be seen because the cameras can’t visualize it. Having the supine extent of the treatment beams on the VMAT arc helps you to position the patient.

We have found this has been really helpful when we’re setting up our palliative patients. And I did mention earlier about the difficult case of the patient scanned with bolus where there was no body for setup; the Beam Guide helped us to set up that patient a lot easier once we switched it on.

Just on this next slide, I’ll make a bit of a caveat that someone forgot to get the screenshot before the Beam Guide was in position. On our single-pulse patients—I don’t know if anybody else experiences this—when you’ve got patients that are quite rotund or barrel-chested, sometimes once you go to 100 at the back of the bed, your ROI sometimes can’t be fully picked up by the cameras. That’s just a patient issue because of the patient size or where the treatment site is. On imaging days, that’s absolutely fine because you’re imaging and you know you’re in the right position. But we did find that sometimes our radiographers felt a little bit twitched on a non-imaging day because they were like, “We trust the SGRT system, but we’re missing this little black hole.” When we got the Beam Guide and switched it on, we found that the hole was in the ROI because you had the Beam Guide, and the treatment field wasn’t necessarily in the whole of the ROI where you would have the white outline of the beam. So if you couldn’t fully visualize your ROI, you used it to set up. As soon as you switched the BeamGuide on, you could see that the orange projection was within the white outline, and you knew your patient was in the right position for treatment. They were more confident once we got this that on non-imaging days we were in the right position for treatment. As I say, spot the mistake: someone hadn’t moved the lung, so the orange isn’t within the white, but that is what I would have liked to have shown you.

As you can see, there’s lots of really good applications of using the Beam Guide for troubleshooting, setting up the patients, and picking up any errors before you actually deliver any radiation.

*This transcript has been AI-generated. Contact us at secretary@sgrt.org if there are any issues. 

Jodie Bailey
Therapeutic Radiographer Team Manager
Norfolk & Norwich University Hospital

So I’ll go through a little bit about Beam Guide and then some case studies that we’ve found it useful on.

We had the system upgraded on the 6th of March. Our previous version was 6.3.2, and we went to version 8.0, so we were an early test site for this release. It was a major release where we had a lot more data and patient information that needed to be transferred across, so it took a two-day process over the weekend.

The main features were Beam Guide and Auto ROI, but as I said, we’ll focus on the Beam Guide part today in the session. We then had more of an update, really, to version—we’re now on version 8.1.1. This was more about listening to the feedback from the early implemented sites and doing some minor software changes.

So Beam Guide, it’s using an overlay of the planned radiation field, so the MLC position. It merges that whilst you’re in Postural Video mode. So it kind of is complementing a setup. It doesn’t replace your imaging; it’s giving you that virtual light field. The huge benefit that we’ve found for other patients is that you can see that exit dose, which you wouldn’t be able to see on a light field necessarily. So there’s a nice case example coming up where that was incredibly useful for us. But yeah, so it’s something to complement setup, and it’s much easier on things.

VMAT plans can be a little bit tricky because you see the whole of the irradiated area. So we’ve found it most useful on things like tangential breast, but it is really helpful to look at chin position on an IMC VMAT plan, too.

How we integrated it within our workflow: we weren’t quite sure how exactly it was going to fit within our workflow, but we use it during the initial setup phase, just checking that’s… and you can toggle it on and off. You can select various fields that you’d like to look at. It can help you make those real-time decisions before you deliver any verification imaging, so it can help support problem-solving prior to imaging. Again, we’ve got some nice examples of that where repeat images and things have been saved by using the Beam Guide.

After acquisition, the white lines still remain at the isocenter position. So for the majority of patients, because we’ve applied a shift based on the soft tissue, we toggle the Beam Guide off. There’s been a couple of patients where we have checked field with the understanding that the patient’s in a slightly different position, but it can give you that gross check.

So I’m just going to press play on here. This is our first example of a breast patient where, during the setup stage, we noticed that the patient had swelling. So I will just pause it when we zoom in slightly. I don’t know if you could see that the swelling was outside of the Postural Video outline. You can see the breast tissue just under here is much clearer, obviously, when you’re actually using it in the room, but you can see on this white line that we were coming much closer to that border. This is something that you can obviously check on your field borders as well, but at the time when you’re making small adjustments to the patient’s position, you don’t want to have to keep moving the gantry to the medial and the lateral field to check. So it speeds that process up slightly, so you can make a decision and then do your final checks if you want to using your actual treatment field light.

The patient did have some roll as well, so we tried to roll the patient slightly to improve position. This patient then did get assessed by the dosimetry team just to make sure the coverage wasn’t impacted, but we wouldn’t have seen that coverage was so tight just based off a verification image. So it does give you that confidence, really, before you start delivering treatment that any swelling or any changes to the breast tissue, that it’s still covered.

This is another really interesting case. We had a palliative breast fungating patient. She had a fungating breast tumor. She came for her CT scan. She was in a VAC bag with limited arm movement at that point. But come day one, when she came to start, how the tumor had progressed, it meant that her arm position was even poorer than it was at the scan. We used Beam Guide because we knew that she couldn’t maintain—she couldn’t put her arms in the same planned position. We used Beam Guide to position the arms to make sure that the hand was out of the treatment field.

We could see the exit dose, which we wouldn’t have been able to see using the field light. We could see there was exit dose on the fingertips, and we were able to reposition her arm to get that out. On this particular fraction, whilst in between us setting up and imaging, she relaxed her hand down. So what you can see in the top corner here is her fingertips from where that had dropped into the treatment field. We saw the image, we saw there was something odd at the top of the image. We then switched Beam Guide on and could see that her fingers were receiving exit dose. Then we used the Beam Guide to reposition her hand out of the way of the exit dose. We didn’t then need to do any repeat images to confirm that it was out of the way, because we were confident that that was demonstrating the position that was needed for treatment.

We then, for this patient, did toggle it on during treatment just to confirm that her hand didn’t drop in again. This is one of the only times where we would use it during treatment. Even though we’d done slight shifts, we were confident at this point for this patient just to confirm that the hand wasn’t in the way of the treatment field.

This saved this patient going back to be replanned because we were able to use SGRT and Beam Guide to optimize her treatment and make sure that we were still delivering it. But it also meant that she wasn’t on the bed for too long, because we could make these decisions with the click of a button rather than having to go in and check field lights. And as I say, you wouldn’t have seen that her fingers were receiving exit dose anyway with the field light.

We had another palliative patient where they were in abdo, but what was scanned all looked exactly the same. It was just like one column almost when you’re looking at Postural Video. Day one, when you have no tattoos, these patients can be quite tricky because you don’t necessarily know where they are on the bed. AlignRT was struggling to detect the patient’s contour; it wasn’t giving them any numbers. It knew the patient wasn’t quite in the right position.

So what the radiographers did is they opened up external beam planning. They could see where the beams were supposed to be going on the patient’s abdomen. They switched on the Beam Guide and could see that they were currently in the wrong position for the patient. So they adjusted the long so that it looked sensible. By this point, SGRT had detected the patient and picked them up, so then they were able to continue on with treatment.

It reduced an error in us setting up to the wrong position and reduced any kind of repeat imaging, because we were able to make these assessments prior to delivering treatment. So it can be a really good value tool to give you that confidence when you’re unsure about your beam geometry and where that dose is going, particularly for those patients where it almost looks like one column and you’re not sure what the long position is.

We found additional applications for clinical value for Beam Guide. One is that you can check your breath hold, and you can visually confirm that your field is covering to make sure that they’re still breathing within that… we’ve got a little margin around. I can’t think what it is off the top of my head right now, sorry.

The other thing is that it gives you that visual feedback to be able to perform corrections on setups before taking that image. It can help improve the patient experience, particularly that patient where it prevented her having a rescan—with a palliative breast patient, where we were able to get her treatment in as soon as possible, and we weren’t having to repeat image. Her setup was incredibly quick compared to what it can be for that group of patients at times.

This is a little video of us doing a setup on a breast patient and using Beam Guide. At the moment, setup with patients in free breathing. You can see a little bit of flash on the chin, so just lifting that up and out of the way. We’re now flicking to the breath hold surface. We’re getting the patient to breathe in. Once we’re in breath hold, we then send everything except overt across.

This is our real-time setup with the use of the Beam Guide. They’re just going to optimize the arm position better there, because you can see there’s a bit of a gap on Postural Video. Getting in one last time just to double check everything, lifting that chin again up out of the way. Then we check FSDs daily and the deformation.

We can see a lot of potential with this technology. What we’re hoping to do—and we’ve had a good thought about how we could use it—is to optimize bolus positioning. Our center uses bolus a lot, so anything we can do to simplify that workflow is going to be to everybody’s benefit. So we’re in the process of thinking about how we could use Beam Guide to help with placement of bolus or checking bolus position during the setup stage.

It’s also about how we can integrate it into our clinical workflows. How do we embed it more within our current processes? Because we’re just getting our heads around it, really, and seeing its clinical uses, but it’s then how do you make it more of an official format for staff?

We’re also expanding our case study library, so the staff will let us know if they’ve found an interesting case where they’ve found it useful or they’ve not, and we can have a look at that. Again, we’re hoping that we can use Beam Guide to strengthen our workflows and improve that consistency, and then reproducibility in setups.

So it’s all about identifying our future development needs and capturing that user feedback to refine how we’re going to use it in the future.