Tracy Reece, BSRT(T)
Oklahoma Proton Center
Helena White, RT(T)
Bryan Hammers, RT(T)
Oklahoma Proton Center
Michael MacFarlane, PhD, DABR Assistant Professor, University of Maryland School of Medicine, Easton, MD
Tags: Accuracy, Workflow optimization, Efficiency, motion management, Implementation, AlignRT
Wendy Tissue, MBA, RT(R)(T)
Mayo Clinic Cancer Center
Wendy Tisue, MBA, RT(R)(T) (00:00):
Hello, my name is Wendy, and today I would like to talk with you all about SGRT and the challenges that we face in protons.
Wendy Tisue, MBA, RT(R)(T) (00:06):
Alright, so, this is an actual one, one of our, Hitachi ProBeam Gantries. We have four half gantry systems. We don’t have any full gantry systems at our clinic. This is what we use, for any anterior tumors that we’re treating, like perinasal sinus brain tumors, that would be close to the surface, but that’s called a bolus helmet. That’s something that was 3D printed for us up at in Rochester. And that has been very valuable for our center as well. But you can actually see our cameras. We have one mounted there over to the side, and then there’s one at the foot end and that those camera positions were very beneficial for our site, especially with the couch rotating all the way around, to both sides with the half gantry. And I have no disclosures.
Wendy Tisue, MBA, RT(R)(T) (00:56):
And, so today’s agenda, I’ll discuss transitioning from C-RAD to AlignRT, designing your proton room treatment sites that we utilize AlignRT with, challenges that we faced going tattoo-less, and then our verification scan challenges.
Wendy Tisue, MBA, RT(R)(T) (01:20):
So our proton center initially opened with having C-RAD as an SGRT system in just one of our four proton vaults. We decided to change to AlignRT due to the difficulties with the C-RAD system and the AlignRT system provided a more friendly user interface for the therapist. This change allowed us to have one vendor department wide since the photon site already had AlignRT. And with having the background C-RAD this did help the proton therapist to be more receptive to learning another new product because you know how excited they always get when they have to learn new products every other year. but since we already had C-RAD in one room, we installed Vision RT in the other three rooms first, and then once those three rooms were up and running, we then swapped out the C-RAD system to AlignRT. And then we initially started, treating our patient, our prostate patients, our breast, and our CSI cases, with, with AlignRT. And then after that we transitioned to open face masks and we are now using AlignRT anytime there’s an opportunity. So meaning anytime that we have a non masked patient will automatically use it for our setup and our monitoring. And I will go into more in depth of our transition from C-RAD to AlignRT and my other presentation tomorrow.
Wendy Tisue, MBA, RT(R)(T) (02:43):
So our room design and half gantry room issues. So these rooms do pose many challenges, but they can be overcome to help find the best solution for your particular clinic. If you do have more than one room, I would highly recommend installing it and commissioning the first room before you go to the other three rooms. and this just helps to ensure that the camera placement’s optimal. And we also did encounter an issue with needing to reinforce our camera brackets. And that was due to the vibration of, you know, these huge gantries rotating. So just have that in mind. But, Vision RT was able to troubleshoot all that for us, got the right brackets in, and we haven’t had any issues with that since.
Wendy Tisue, MBA, RT(R)(T) (03:26):
Another unique thing that our center has is apparently we were the first system to ask for three monitors for each room. So we have the monitors, actually one outside in the control room, and then we have a monitor in the treatment room, like where you’re actually setting your patient up. But then we also have an in-room imaging suite so that we can, you know, pop behind the wall, do all of our imaging, and then we go out. So we actually have one of those in three. We have one in each of the three areas. and that’s provided us with the ability to easily see and recapture the patient at any time needed. Right here our therapists use like some translucent paper and then they actually put cheaters on. It says like right left in soup table down and up, chin up, chin down, roll your left side up, roll your right side up hips, right hips left. And that’s because we can’t send the coordinates to the couch. We have to move our patient rather than the system correcting the table for us. this helps the therapists kind of with that anxiety feeling like, oh my God, I see these red bars going to either side, but I don’t know how to correct for it. Which way do I move my patient? So this was just like a quick little cheater for our therapist. Some don’t need it any longer. They’ve took, they’ve taken those cheaters off other people, they like them, and as they rotate onto their machines, they’ll stick them back up. So whatever helps your therapist be more the most comfortable, that’s what I would say do to to help them.
Wendy Tisue, MBA, RT(R)(T) (05:01):
So one of the initial struggles, that staff encountered when starting to use SGRT was when there was a discrepancy between the SGRT system and imaging. If they encounter this, they have been instructed that imaging will always take precedence over the SGRT system, but we did need to understand why there was a discrepancy and was there a solution to it. So they learned that the way an ROI is drawn can, change how well the SGRT system aligns. So the pelvis tends to be one of those tricky sites, that they’ve gone back and forth with the ROI shape. especially with having that little strip across the belly. we found that sometimes when patients are larger, that strip can be a hindrance more than a help. but it can also be helpful, especially if you have a half gantry and you have to rotate the couch back and forth. So some of our larger patients, we do have to keep it off, but some of them we do like to keep it on just so we don’t have that camera flickering issue. But we do also ensure to exclude the buttocks, and then also going too low into the leg, because that can introduce some unnecessary pitch issues that aren’t really there. And then you notice that when you take your x-rays.
Wendy Tisue, MBA, RT(R)(T) (06:16):
Here’s an example of a prostate ROI. You can see in this example that we did not include the belly strip. so our prostate setups, include a CIVCO, leg immobilizer that’s always indexed, in a standard position for all patients. Our center is very big on indexing every patient. I can’t think of any site that the patient’s just laying on the table without being indexed in some manner. This eliminates the need to make vac locks, it saves, storage space. It eliminates the issue of having the bags deflate, which if you use vac locks, we’ve all had that at one point or another where you put a device into storage, it deflates on you and then you’re either calling the patient back for a recent or fixing it on the first day. So we did exclude the belly strip on this example. And then, like I said earlier, we avoid going too low on the leg, for the pitch issues. And then once again, we excluded the buttocks and then we also make sure that we do not have any bunched up jeans or pants underneath the patient. I know some centers they don’t have their patients change a lot and just have them kind of lift up their shirts or lower their pants, but then you have all this material bunched up under you. And especially in protons, we don’t have the luxury of having some something introduced, into the treatment area. We need it to be as precise as possible. so we’re, we’re pretty strict on that. If the patients don’t wanna change into the scrub pants that we have at work, we just instructed them to wear, you know, like pajama pants or basketball shorts, something like that that they can easily lower. That won’t be all bunchy but no belts.
Wendy Tisue, MBA, RT(R)(T) (08:00):
Here is one of our, breast ROIs that we drew. So our breast set up consists of the Orfit system. We do have the AIO system for orit, but for this particular setup they just have it’s like a little black breastplate that we put on the table with a lock bar and then we utilize the back lock and then we also use the orit knee cushion. And then we also document which risk the patient is holding and then we have them turn their head. So you can see that we do, scan pretty high for CT sims. Somebody did ask a question in an earlier talk, like, how high do you guys scan for your, your breast patients? We, we like to go high for our dosimetrist, but then it also helps to provide a nice, a nice image for the therapist to utilize for the postural alignment to know exactly where those elbows supposed to be the head turned, is that correct? So it shows which wrist we have their head turned and then variant or vision has allowed us to ensure that the arms and chin are always in the correct position. And then many of our breasts are reconstructed breast patients. So we don’t have a lot of the pendulum breasts over in protons, but when we do, sometimes we’ll exclude the breast tissue and draw more of a ROI across the clavicles and the sternum. And that’s helped sometimes with, you know, the breast tissue. another thing to note though, with protons and breasts, we do, when we first started we had a mask set up, but then since utilizing vision that allowed us to go mask list for a setup, the exclusion for that is when we treat bilateral breasts, we do still utilize a mask system just because you don’t wanna graze into the other breast at all and the position for the actual breast tissue is very critical.
Wendy Tisue, MBA, RT(R)(T) (09:55):
CSI and SGRT, one of our biggest challenges with CSI setups was the alignment process. We see a lot of CSI treatments and this is a site that is not easy to image or treat and we do have a great setup for CSI by utilizing a body fix bag in the CDR head frame and mask. And the next part that was needed to improve upon was the imaging struggles. So before SGRT, we were imaging the brain then all the way down the spine to ensure that the body was perfect before we started to treat. And then SGRT has allowed us to reduce some of the initial imaging at table 270 and table 270 for us is our home position. That would kind of be like a linac table zero. And this is where we always do our initial imaging, with our 45 degree like images.
Wendy Tisue, MBA, RT(R)(T) (10:46):
So we don’t have a nice pretty AP and a lateral, we have 45 degree images. So with SGRT we were able to go from imaging the brain and all the way down the spine to just imaging the brain and then ensuring that the body was aligned with SGRT before we kicked the couch over to that first treatment angle. And once we kicked the couch to the first angle, we then imaged and treated the brain. After we treated the brain, we would then shift down to the spine, imaged there this time and make sure everything looked perfect and then we would proceed with treating the spine. And by doing this method, it saved us about 15 to 20 minutes of extra imaging, and going, between the isocenter. Okay, so it saved us about 15 to 20 minutes of extra imaging and going between isocenter in the beginning, but on the first day we will still image, the lower spine at table two 70 just to ensure that between imaging and AlignRT, that there’s no big deviations, that it all makes sense. And then when we draw the ROIs for csis, we only use one of the spine isocenter. We do not set up ROIs for each of the ISOs in vision, since we’re not typically using vision to check the head position. That’s another reason why we don’t do the head one. The photo on the left is of a pediatric patient and the therapist drew the ROI across the chest and then the mids spine and then the hips and then the adult patient on the right, he had a larger belly so they drew it across the chest again and then used the hips, kind of like prostate ROI. And since we only use the one ISO on AlignRT once we start treating and then we shift down to the lower spine iso, we’ll actually recapture for the day and then use that for monitoring only. And then we shift to the next ISO and then do it again. So CSI are kind of a complex one. Emily was great with helping us figure out a workflow. You know, do we actually wanna switch between all the three different ISOs or do we just wanna use it more for making sure your spine’s nice and straight. The question was do we do spine imaging at all? So on the first day we’ll image the brain and then we’ll align with AlignRT and then we’ll still image the spine on that first day. But once we know that yes vision, yes the x-rays agree with each other on the subsequent days, we’ll image the brain, make sure that the spine’s straight with vision and then we don’t spend another 10, 15 minutes imaging at table 270. We’ll just kick right over to our first treatment angle, start treating with the brain. But then when we shift onto to the spine ISO there, we still image there. And our physicists, they’ve created something called a bounding box for our CSI. Josh Stoker did that. and so that has a built in kind of tolerance that as long as that spine is within this bounding box that we know that we’re treating to the proper dose. So then that way we’re not fuzzing with like the super small millimeter shifts. just if as long as that spine’s in the box, then we’re good to treat. So that’s really helped the process as well.
Speaker 2 (14:15):
Is that the most included process?
Wendy Tisue, MBA, RT(R)(T) (14:19):
Yes. Josh, if you guys have any questions with the CSI planning Josh Stoker, I could get the contact information for you and I’m, he’d be happy to share that.
Wendy Tisue, MBA, RT(R)(T) (14:31):
Yep, yep. So we’re just using the one, spine ISO and then we’re able to use this for either ISO
Speaker 4 (14:42):
And do you ever have an issue when you’re using ISO cameras?
Wendy Tisue, MBA, RT(R)(T) (14:49):
So what they’ll do, so say this is like the soup spine ISO, for that larger patient, if they have a second spine ISO, we’ll you know, go to the treatment parameters of whichever one it’s drawn on first. So say it’s the soup iso, we know it looks great there, we x-ray and then we capture for the next day only and then we know when we go to the next ISO center that it’s gonna look off. So we just automatically recapture it again. But then we’re still always imaging with x-rays to make sure that it looks good. So essentially we just use this on the very first spine iso to know that your patient’s body is straight. We, we just kind of trick the system to let us keep going.
Wendy Tisue, MBA, RT(R)(T) (15:34):
So open face masks, they’re our latest adventure at our proton site. So our boss masks, they’re not a fan favorite, due to the inability for the therapist to see through them the rigidness of the masks and the difficulty with fastening, to the ball board, but they do their job well and we do have a very low replan rate for head and neck and brain cases. So when we asked about changing the setup from the closed face mask to the open face mask, our physics and our physician group were very hesitant. So we proposed doing an evaluation of the two open face mask types and compare the findings against the closed face masks with the similar disease sites. So we had a total of 10 open face, head and neck masks and 11 of the short masks. And we did a retrospective analysis of the number of re-setups, number of replants and the setup issues. We also looked at image guidance shift information exported from Aria and this was analyzed in Excel for, for statistical significance using the student’s T-test. And the calculations and assumptions were peer reviewed by physics and they found that they only had a lateral shift, on the open versus closed long mask that was found to be statistically significant. And the difference was a p value of a whole whopping 0.02. So their interpretation of the data is that the mobilization is not compromised by using the open faced version of the mask. And after the evaluation was complete, the physician group agreed to transition to the open face mask with the exclusion of perinasal sinus and their cord and the chords. And that’s also primarily because we would use that bolus helmet that I showed you in that first screen. So the bolus helmet is essentially covering everything that vision could see. So we just wanna make sure that the patient would have less of a chance to move. So if we’re using the bolus helmet, we would use the closed face mask, and not utilize vision for that for any of our open face masks. they’re not utilizing the bolus helmet, so we use vision for setup and for monitoring. So with the new setup, our, so with the new open phase setup, we did find that our patients are less anxious. The setup times have decreased, due to less resets as well. And in the beginning we did have some challenges with the system saying that we needed a pitch. but when we would x-ray the, the x-rays would find the opposite. So we found out from the trainers that we were not stretching the opening of the, open face mask down low enough to the anterior part of the ear, or we were having the ROIs actually go too low and they were touching the mask, which would cause inaccurate deviations. so by fixing those two issues, that has helped, with those pitch issues. right now our biggest challenge is noticing that there is some gapping in the mask at the forehead and this may be due to using a shim when making the mask. So we’re in the process right now of trailing no shims on five patients to see if we notice an improvement at all.
Wendy Tisue, MBA, RT(R)(T) (18:45):
We are now using a AlignRT for any non masked setup, in protons. And the pediatric patient on the top was having their whole right lung treated and Vision RT was very helpful with allowing us to avoid tattooing the patient for these treatments. And the patient on the lower left was also a lung patient. Typically we have adult patients in an orphan mass to help limit, the chest movement and reduce breathing motion as well. But she did have an allergic reaction to the mask, so we had to modify her setup to a vac lock wing board setup, with just a knee cushion and we were able to use the postural alignment to ensure her chin was also in the proper position since it was a higher lesion as well as to help position her body daily and monitor her for movement. And so far no issue. So we’re hoping maybe we could convince physics and the physicians to maybe go to more of a massless setup, especially when we know that there is just one cm or less of that tumor motion. The patient on the lower right was having their right thigh sarcoma treated and you may notice the bolus on the opposite leg and the genitals, this was to help protect his anatomy from getting any dosed, due to the ankles of the beams needed for his treatment. And we were able to use AlignRT to help not only with the positioning of his leg, but to also help with the positioning of the bolus.
Wendy Tisue, MBA, RT(R)(T) (20:12):
Tattoo list challenges. So being tattoo list is amazing, but it can have its challenges, especially for protons. if the system is down for treatment for our breast cases, our sim does have kind of a protocol that they do for our breast and our prostate cases to help, document some distances. So sim will actually document the distance from the sternal notch to the CT origin. And CT origin is just where we drop our, our treatment ISO, and dosimetry will provide any per plan shifts from there. But due to having robotic couches, we do utilize an Excel sheet that helps us calculate the table coordinates for the patient’s setup, and these calculations are actually inputted into, treatment prep before the patient starts treatment. So in the instance AlignRT is down, we’ll just visually straighten the patient, ensure the correct hand is holding the correct wrist, and then go to our saved table coordinates and then just start our x-rays from there. Our prostate patients, sim once again, they’ll do umbilicus table coordinate at time of sim and they’ll also take pictures. So if vision would be down, we would visually straighten the patient, use that umbilicus table, coordinate for the longitudinal position center, the patient go to the table coordinates, which are once again calculated by the DOSSI and the RTT at time of chart check and pre put in. So we haven’t had any issues with that, but honestly I think since we’ve been up with vision, it’s only been down maybe twice on us and it wasn’t system wide, it was just in one particular room and it, there was an issue with our, our network. Nothing with vision,
Wendy Tisue, MBA, RT(R)(T) (21:58):
Many treatment sites do require multiple CT verification scans to check for anatomy and tumor changes. we do typically do the CT scans, in our room that has a CT on rails, but not all of our rooms have a CT on rails. So, and since we don’t have cone beam CT, we only have our lovely KV imaging, which is another reason for having the one CT on rails. So in the instance that a patient needs to have a vSim, we will do them in in gantry one with the CT on rails, but if the treatment site is too low, we’ll have to take them into one of our sim rooms. Or if a patient also needs a four DV sim, since we don’t do those in the treatment room as as well, so if it’s too low such as a CSI, or something really low on the pelvis, we’ll take them into sim and do the V sims there.
Wendy Tisue, MBA, RT(R)(T) (22:57):
But since we don’t have an AlignRT system in sim, that does make it a little bit more challenging or if the patient, needs to come for a vSIM before the patient starts. But if we do go from treatment room to CT sim, we make sure that the therapist know to add those, setup marks onto the patient. So then that way the sim therapists have a way to align the patients. So Vision RT please make some sort of a SIM system for us for US proton Centers. and then we do have the SOPs, to account for all of the situations if visions would be down.
Speaker 6 (23:37):
So how, how often do you guys do that?
Wendy Tisue, MBA, RT(R)(T) (23:42):
So a lot of our head and neck patients, we have V sims done weekly. but those are typically all done in gantry. One where we have the CT on rails, a lung patient, or like a lower esophagus or liver, something like that where you need that 40 scan. Those will be done in CT sim. And that’s just because our RPM system is set up for treatment mode, not sim mode. we don’t currently use, Vision RT for breath hold or gating just because we do both of those treatments. And I believe at this time vision can only do the DI having RPM that can do both of those breath hold or the gating treatments. That’s why we’re not currently using vision for the breath hold, but if they would get that, we’d switch.
Speaker 2 (24:35):
Once you bring the we the plan to the new evaluation ct that plan by the dosimetry that you’re looking for, how do you go?
Wendy Tisue, MBA, RT(R)(T) (24:51):
We do a replan. So, we don’t have a lot of replan. The plans are nice, they’re robust, but if they ever find that something isn’t within, their parameters or the guidelines and they will do a replan. Just to hit on photon, patients typically don’t have this issue because they’re not doing all the V sims like what we’re doing in protons. But in conclusion, AlignRT has proven to be an invaluable resource in our efforts. There are a few challenges that we have encountered along the way, but they’re all manageable and can be overcome with the right strategies and dedication. The benefits and positive impact of AlignRT far outweigh these challenges, making it a crucial component of our success.
Abdallah Quballa – SGRT Europe 2021
Tags: Accuracy, Workflow optimization, Efficiency, motion management, Implementation, AlignRT, proton
Erin McCabe, RT(T)
Proton Radiation Therapist
Mayo Clinc, USA
So, some learning objectives. We’re going to do an overview of our proton therapy center, and then we’re going to talk about the role SGRT plays in ensuring accurate patient setups in proton beam therapy. And then specifically, we’re going to talk about our breast workflow. Then we’re going to talk about some challenges and how we overcome those. And then we’re going to talk about some benefits, and then summary at the end. Let’s get started.
So, some background information, some key terms, and knowledge to get you all informed and understanding. So we have 180 degree Hitachi ProBeam gantries. We have four different treatment rooms and one proton beam. We have a six degrees of freedom couch, which you can see here is attached to a robotic arm. This is the gantry. This is where the protons come out of. We have our X-ray panels here, and we have our X-ray tubes built into the floor, and so we have oblique X-ray kV imaging.
Over here, we have our CT scanner. So what we can do is called CT onrails, which is where we move the treatment table over to the CT scanner, do a quick CT scan, and then move the patient back over, and we can realign them using our planning CT scan and the CT scan from today.
So, something to note, we do still use tattoos. I know, it’s a hot topic today, and we hopefully will soon be moving away from that. We do use UV tattoos, which are pink in color, and we have a black light that our therapists use daily to help line up the patients. They are less noticeable, but nonetheless, they are a tattoo. The reason that we do this is because not all of our patients that are simmed for protons actually get proton treatment. You guys probably know with insurance, all that lovely things. So because our patients can potentially go photons or to our regional sites, we do have to tattoo them in order for those other sites to treat the patient because those sites do not utilize SGRT, unfortunately.
Patient immobilization. Our breast patients are positioned on an index breast board with a chin mask in place and a small back lock used for immobilization. So whichever side of the breast that we are treating, that arm is raised up above their head holding a handle, and the opposite arm is at their side holding what we call a slhandle or a slanted handle. These are all going to be indexed to the table for consistent positioning, and it’s helpful for us once we get into our SGRT workflow. Something to note here is that our bilateral breast patients are immobilized in a five-point mask. So what’s important here is that we do not include the ROI when importing SGRT to cover the part that is above the mask. So we only want to encompass the tissue that is below.
Fractionation. So what I’m going to consider our standard breast will be our press chest wall with or without nodes. This will be 15 fractions, and those patients will have a chin mask and a back lock for immobilization. Like I said, our bilateral breasts, these are going to have a five-point mask. And then our partial breast PBIs, these are going to be three to five fractions, but something to note is they will not have any patient-specific immobilization. They will still be on the breast board like our standard breast, but without that patient-specific immobilization.
So some key terms here. We have Setup 270. This field is used to establish the patient’s treatment position. This is where we take our X-rays, where we get the patient on the table, kind of loaded up and everything looking perfect before we begin treatment. So 270 is referring to the table angle. Like I said, we only have 180 degree gantries, so we have to move the table to different treatment angles.
So then CT Sim/DICOM Body Only. This is referring to the patient’s external anatomy. This comes from the treatment planning scan, and we will see that in a few moments here. And then the ROI, region of interest. This is created based off the body only, and this will encompass the treatment area.
So why is SGRT useful? So this helps us make sure that the patient is lined up, and where we have lined them up post X-ray alignment, they will stay in a certain threshold before delivering treatment, and especially during couch rotations. So this will monitor the patient’s position relative to isocenter at each treatment angle and ensuring rotations like table 180 to maintain isocenter alignment within four millimeters translational and three degrees for robotics. And again, this three camera infrared system provides continuous surface tracking during treatment and detects drift from isocenter instantly. And especially before beam delivery with having four treatment rooms, we do have longer beam wait times, and so making sure that the patient has not moved during those long beam waits is crucial for us to have an accurate and successful treatment. Overall, these principles show that we use SGRT to enhance confidence before and during beam delivery and ensuring that we will have the most accurate treatments.
So, this is our standard screen. We’re going to do a little overview in parts of the tools that we use daily for breast treatments. So up here, where we have this little person icon, we’ll have the patient’s name and date of birth, and potentially their MRN number. We will have the play button here, then we have the treatment plan. So you can see this patient has had two replans. This is the initial planning scan. We call it F1, which stands for fraction one, breast left. So each replan a patient has will create a different CT Sim DICOM body-only structure because, again, it’s based off that initial CT scan.
Then over here, we have the CT Sim body only. That is going to be the purple pixely outline of the patient. And F19 breast, this is our ROI name. So this is coming from the plan that they are on, and this is going to be the white structure here. To the right, we have our vert, long, lat, magnitude, yaw, roll, and pitch. Our translationals are going to include our vert, long, and lat, and then our robotics will be our roll, pitch, and yaw. Down below, we have our surface deformation and video, which we saw earlier during those presentations. Reference is for post x-ray alignment. Treatment is comparing our patient’s current surface to that treatment planning scan, and then couch 270 is referring to the table angle that the patient is at.
So how do we implement this into our daily breast standard of practice? So this is all before x-ray alignment. I just want to emphasize that. This is going to be whenever we get the patient on the table at setup 270. What do we do? How do we do it? Why do we do it?
So initial setup, we’re going to start with that setup 270 field, and we are going to align the patient using tattoos and room lasers. Then we are going to shift the couch to its planned iso-center. So those are from our dosimetry team and our patient’s information note. Then we’re going to use the SROT surface capture. We are going to select the treatment button. Then this will compare the live surface to the reference CT Sim DICOM body only. So now we’re going to evaluate what it looks like. So for translation tolerance, we have four millimeters. In robotics, we have three degrees. Then we’re going to apply these corrections and therefore manually adjust the patient if needed, if it’s over three degrees. Then after we’ve made these adjustments, we will do another treatment capture, which will then hopefully all look fabulous. Then we are able to step around the corner and take x-rays.
From the last talk, we talked about the postural video, so we use that as well. This helps us before taking x-rays to hopefully help aid with visualizing the arm positioning, because that can be a bit of a struggle. Even though the patient is in a vac lock, this can still be helpful in helping consistency of arm placement during setup, and it reduces time by minimizing the need for repositioning. So as you can see here, we have the three different cameras, and you can see the different longitudinals or intra-post, you name it.
So post x-ray alignment. So our x-ray alignment, we are lining up to the chest wall, sternum, anterior ribs. We’re looking at the arm. We’re looking at potentially the chin to make sure that eye is up and out of the way for treatment. So we have the process on your right and the purpose on the left. Our auto-selected ROI is going to be ensuring that the correct treatment plan is selected. Then we’re going to enable that play button, which is going to activate real-time surface tracking. Then we will select deformation, which was at the bottom next to surface and video. So the deformation is going to give us a percentage. So this is going to let us know, because we cannot see soft tissue on our x-rays, what does that actually look like compared to the actual treatment plan that the physician planned. Then we’re going to analyze this. We can see some different colors. We can see green, we can see red, we can see blue. We’ll get into that a little bit later, okay?
So then we’re going to view the results, identify the magnitude, and the location of the surface deviation. Sometimes we can run into troubles, which we’ll get into later, but we will also report screenshot. So this is big because it gives our physicians the ability to see what exactly we are treating on a day-to-day basis. So this is super helpful if the patient is getting a replan and the physician’s like, “Hey, could I possibly pull up these documents?” And we can easily have them at their reach at any moment. So the physician could evaluate for a replan and decide if the deviations are consistent enough to justify a replan. So seeing if this is a daily issue or whatnot.
So this is going to be an example of a report screenshot. So this is exactly what the physician will see whenever they ask for those documents or pull up that file for the patient. So again, the patient’s name would be right here. You can see that the play button is pushed here, which is why it’s now a pause. We have the plan name, we have the ROI. We can see that these are all within our tolerance. They are within the green. Again, our translationals are going to be four millimeters, and our robotics are going to be three degrees.
So basically, whenever we press deformation, this little box will pop up. This will show what we’re mainly focused on is the percentage here. We obviously want that closer to 100 as possible, but we understand that over 85% will still be an accurate treatment. And something to note here is that this is a standard breast. So this patient’s currently on fraction 19, and the tolerance limits for this patient’s treatment are going to be above and below five centimeters. So you can see this is all green. Fabulous. Then we’re going to come over here to the corner and we’re going to report screenshot. So just by pressing that button, it will automatically save to our database for our physician’s review.
Now, this is an example of a PBI. The key thing to note here is that everything is the same over here. We have the patient’s treatment plan. We have the name of the ROI. We’ve pressed the treatment button. Now we are looking at the deformation. Again, this top percentage is at 90%, which is pretty great. And you’ll notice here the tolerance limits. So these are going to be above and below three millimeters. So this is adjustable for us to move based off of what the patient’s treatment is. And as you can see, the CT sim body only is kind of adjusted. So we can do that because it is a 3D structure when it is imported. So we can get a better view, especially if there’s any color mapping or whatnot. We can get a better view and see where that is. And again, you would report screenshot, and it would go to the physician for treatment.
So some problem-solving color mapping. As you can see here, we have some red and some green, and you can see our VER is out of our tolerance. And in the top corner here, our percentage is looking around 54%. So that is not treatment approval okay. So what we can do from here– So let’s explain a few things before we get too excited. I’m excited.
Red means there’s too much tissue. Blue means that there is a lack of tissue. So concerns a large area of red or blue within the ROI suggests that the breast position is not in the same position. So striations can also occur. What is striation? It’s almost like a lightning bolt within the ROI. So what you can see is almost like a zebra print of red, blue, red, blue, red, blue within the breast. So this is very concerning, and the physician will most likely be called because of this. So we do not take this lightly whenever we are analyzing these before treatment.
So we’re going to have issues because of a flip magnet. Magnets are used for our breast patients with implants to help the nurses or whoever is filling those with solution, help locate that, the breast position differing, or sometimes just patient movement. So normally what our go-to steps are, if the percentage is below 85%, we will– So the steps here, we press play, we’ll analyze. We’ll see that the stats are below our 85% threshold percentage. Then we’ll take an additional X-ray. Sometimes the patient just moves, and that’s okay because we can realign and get them back into the right spot, or we can potentially call physics for guidance. So our physicist would pull up the plan and external beam and analyze the discrepancies within the ROI to see as if they feel it is significant for their plan, or we can just continue on like normal. The physics team will also make a note of this, and so will we in our treatment log to journal about this, so whoever is treating the patient next could know beforehand.
So how is SGRT beneficial for patients and therapists? For patients, there is no additional dose during setup. So we do still take X-rays post initial setup, but this hopefully will minimize that by getting us closer to that isocenter alignment, and it’s improved patient positioning. So actively comparing the patient’s current surface from today to that initial planning scan will hopefully give us that perfect, most accurate alignment that we can, and it lets us see better soft tissue alignment. Again, we cannot see soft tissue on our X-rays, unfortunately. But this gives us a window into, okay, if we did, this is what it would look like. More consistent arm and body positioning, again, using the postural video that helps us aid in positioning an arm for treatment, and then reduced risk of setup errors. So sometimes with our breast patients, they are indexed differently on the table on whether or not they’re photons or proton simmed. So sometimes getting those gross errors out of the way or realizing before we take any X-rays that we need to adjust something, we can easily without giving them any extra radiation.
And for therapists, so this allows us to set up without solely relying on X-ray alignment. So it gives me peace of mind, especially coming from a photon background of being like, “What? We can’t see this.” And it’s like, no, we have this software that we can use that will give our therapists confidence and allow them to see what exactly we’re treating a little bit better. So then we can see the patient from multiple angles simultaneously, while also this is used pre- and post-X-ray alignment. We use this also to monitor the patient during treatment. So after we’ve analyzed the deformation, after we’ve reported screenshot, then we will take a reference, which will then capture the patient’s current surface and make sure that they are perfectly aligned for treatment. Also with the postural video, we could use that for checking clearance. So even though we are checking clearance with the table, patient, and the gantry while we are in the treatment room, it’s also helpful to give you peace of mind if they were to move because yes, the breast that we are treating is the ROI, and that’s what we are most focused on making sure is not moving during treatment, which leaves us room for error when it comes to the patient’s arm. So they could move their arm whenever we are outside of the treatment room and us not necessarily catch that. But we can turn on our postural video during treatment and make sure to see if the patient has moved or not. And then it supports efficient and reproducible setups. So again, using the CT sim DICOM body only, we are coming directly from the patient’s treatment planning scan and today’s current anatomy and comparing those and making sure that those line up perfectly.
So to summarize, SGRT is used prior to X-ray alignment using the treatment button, which would compare the patient’s current anatomy to that initial CT planning scan. Then it’s also used post X-ray alignment by pressing the play button deformation and analyzing the patient’s surface. Again, we are going to adjust the tolerances here to either 0.5 or 0.3, depending if they are a standard breast or PBI. And acceptable percentages for treatment are 85%. However, this is up to physics or the physician’s discrepancies. And SGRT is used during the patient’s entire treatment to monitor movement. This is especially helpful during long beam waits. And if the surface capture is out of the four-millimeter translational or the three-degree robotic tolerance, we can easily pause the beam from outside of the room and take an additional X-ray and realign. If we do realign, we do check the deformation and take another reference capture for the remainder of treatment.
Mindy Joo, PhD, DABR
Medical Physicist
Inova Health, USA
Lei Hu, PhD, DABR
Medical Physicist
Inova Health, USA
Transcript
The objective of this presentation is to understand the challenges of implementing SGRT in proton therapy, to explore the impact of ROI design on SGRT in proton therapy, and to evaluate the effectiveness of extended ROIs in maintaining stable and reproducible service tracking.
A little bit about Inova. We’re a health system right outside of DC in Northern Virginia. We have several hospital sites spread out across the region. Our Department of Radiation Oncology is part of the Schar Cancer Institute, and we’re located in five different locations. The Fairfax location is our main medical center, and it has some of the special machines like Proton and CyberKnife. But across the system, we have about 11 to 12 LINACs, and each LINAC is equipped with AlignRT system, including C-arm and ring-based.
So yeah, talk about the Inova Schar Cancer Institute and the proton therapy in Inova. So yeah, Inova Cancer Center proton therapy is located in the main campus, the Schar Cancer Center. Okay. So there we have a IBA Proteus Plus system, which contains two rotation gantry treatment rooms. So the two rooms are identical so that patients can be transferred freely from one room to the other. So the system has a pencil beam scanning with small spot size, and then we have a six-degree robotic arm couch. Both rooms have dual kV and a CBCT OBI system. And then the treatment field is 30 by 40, it’s a large field. For the breath-hold, we use SDX system. We have two AlignRT systems in both of our rooms for setup and the monitoring of the patients.
Okay. So the Inova Proton Therapy Center started in March 2020. Over the past several years, we have treated now about 1,500 patients. So among the patient size, 30% of them are head and neck patients, and then we have about one-fourth of prostate and pelvis patients, which is divided half by half, prostate and pelvis. Then we have about 15% of chest wall and breast patients, about 10% of brain patients, and we do have a small amount, 2% CSI patients. Then we also have the thoracic patients and the lung patients, about 6%. And then abdomen patients and liver patients. Yeah. Basically all sites.
We’ve been active SGRT users for several years now, both on the photon and proton sites. SGRT has been fully integrated into our clinical workflow and plays a key role in our day-to-day patient setup. Our proton workflow is pretty similar to the photon one. We’ll first do an initial setup using SGRT once the patient gets on the treatment couch. For every proton patient, they always do a cone beam CT afterwards. And based on the cone beam CT, they’ll do fine adjustments before verification. And once that is done, we’ll capture a new surface reference on AlignRT, and we’ll continuously monitor the patient during treatment. SGRT has allowed us to improve our efficiency in patient setup. Our RTTs are very reliant on it. That’s the first thing that they go to once the patients are in the room on the couch. It also allows us to eliminate skin marks and tattoos. Several of our sites have gone tattoo-less and marker-less now. So eventually, we’re going to try to do that across the entire system. SGRT also streamlines our clinical workflow overall.
Proton therapy has a unique challenge that’s very different from photon. It has gantry-induced occlusion of the SGRT camera’s field of view. So for example, for left-sided breast patient, based on the treatment plan, the gantry and the snout might block the SGRT cameras completely. So if you look at this photo examples, so this left-sided patient, if you look at the AlignRT, it’s not getting enough information due to the gantry blockage. So all we can get data from is just inferior to the breast tissues, and that’s usually what the ROI we draw for each patient. That’s based on where the treatment site is. And proton, again, based on how the plan is done, it could block a significant amount of the signals from the camera. So when that happens, it reduces the surface tracking continuity, also the accuracy in treatment monitoring. And because of that, that might cause interruption to treatment delivery, and we might have to go back to cone beam CT and increases our reliance on image guidance. And might also prolong our treatment duration overall.
So in that case, we decided to investigate using an extended region of interest for some of our proton patients. So in this study, we expanded the ROIs to include surfaces adjacent to the treatment site. So for a left-sided breast patient, we might extend it over to the right side as well. And if you look at those photo examples, without the gantry in the way, the cameras are able to see the entire ROI. And then during treatment, even though the gantry and the snout are blocking some of the cameras, the cameras can still see the other side of the ROI. So that will give us enough information to continue the treatment and ensure that patient’s in the same position, is not moving.
So in this study, we divided our data into four patient groups. We looked at thorax, pelvis, prostate, and breast patients. So we used each of their log files. After each fraction, we extracted those data from those log files and analyzed them for each patient.
Okay. So the log file from VMRT actually can give us a full recording throughout the treatment. Here we show the deviation of the magnitude throughout recording. This example. From here, basically we can clearly see the patient that went through the whole process during the treatment, starting with the initial setup with AlignRT. Then we proceeded with the CBCT to adjust the patient position. After that, we do a rebaseline, and then here we can monitor the treatment throughout the course. And if we take a closer look at the recordings during the beam on, you can see. This is an example of 200 seconds during the treatments. You can see that the oscillations of the deviations, and then basically here you see two trends. One is that you see a little bit of drift throughout the course, but it’s still within two millimeters, basically. And another is that the curves oscillated. This reflected the breathing of the patient.
So we can do analysis on these recordings and then for different sites and for different patients. Here we show the results. For each site, starting with the breast and chest wall patients. And here we show three patients’ data just to see the variations among different patients, and also that taken in different days, from day one to day five in consecutive days. Each data point here is the mean value of the deviation with the standard deviation within that fraction. And also across five different fractions, we can get an average of the mean deviation and also standard error of the mean for that mean deviation. We can see from among different patients that the mean deviation is about one or two millimeters for the five days. So this is for the breast patients. So with this, we can see the intra-beam reproducibility and also inter-fraction reproducibility.
So this shows the result for the thorax patients, also three patients. We can get the mean and standard deviation for each fraction, and also that across the five consecutive days. And we can get the inter-fraction reproducibility with the mean value and the standard error of mean. And here we can actually compare this with the breast patients. We can see a little bit increase of the inter-fraction, the mean deviation. Seems that the thorax patients have a little bit larger deviation compared to the breast patients. Also, here we show the pelvic patients across the five days, and then each day of the mean and standard deviation for three patients. And we get similarly, the mean value and the standard error of the means among the five days. And also the same thing for the prostate patients.
From our study, we found that SGRT provides precise patient positioning and real-time treatment monitoring without compromising workflow efficiency. And with proton therapy, gantry-induced occlusion of the SGRT camera is a major challenge. However, extended ROIs might help with establishing stability and reproducibility intra and inter-fractionally. In the future, we will explore the application of SGRT for beam gating in proton therapy, for example, for DIBH breast patients or any thorax patient, and evaluate its performance compared to other breath-hold methods like ABC. We currently use SDX for all the breath-hold patients.
*This transcript has been AI-generated. Contact us at secretary@sgrt.org if there are any issues.
Wendy Tisue, MBA, RTT & Sarah Carroll, RTT
Mayo Clinic, AZ, USA
Tags: Accuracy, Workflow optimization, Efficiency, motion management, proton, Implementation, AlignRT